6.4.6 Speed Calibration
Keep the belt weigher running at a constant speed. After the belt running time of a revolution being
measured by stopwatch, do this operation for correcting Speed Coefficient.
0010.00 [m]
R: 0.01-5000.00
P102 Belt Length
Main Display Interface
【MENU】+【▲】【 ▼】: F2 Calibration
【ENT】+【▲】【 ▼】: 6 Speed Cal.
【ENT】
M
Run-timer over
New: 100.0pl/m
Pulse: 980
6 Speed Cal.
Old: 100.0pl/m
【MENU】: Exit 【ENT】: Save
M
[P103]: Speed Cofficient [pulse/m]
【ENT】
0028.0 [s]
R: 1.0-9999.9
Run-timer
Length: 10.00m
Timer: 27.0s
6 Speed Cal.
Pulse: 35
【ENT】
Pulse Count
Belt Length Set value