EasyManua.ls Logo

Supmeter BST100-E81 - 6.4.7 Belt Length Calibration

Default Icon
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
43
6.4.6 Speed Calibration
Keep the belt weigher running at a constant speed. After the belt running time of a revolution being
measured by stopwatch, do this operation for correcting Speed Coefficient.
0010.00 [m]
R: 0.01-5000.00
P102 Belt Length
Main Display Interface
MENU+】【 : F2 Calibration
ENT+】【 : 6 Speed Cal.
ENT
M
Run-timer over
New: 100.0pl/m
Pulse: 980
6 Speed Cal.
Old: 100.0pl/m
MENU: Exit ENT: Save
M
[P103]: Speed Cofficient [pulse/m]
ENT
0028.0 [s]
R: 1.0-9999.9
Run-timer
Length: 10.00m
Timer: 27.0s
6 Speed Cal.
Pulse: 35
ENT
Pulse Count
Belt Length Set value

Related product manuals