AMI Oxytrace QED
Installation
A-96.250.531 / 300617 25
3.5. Electrical Connections
WARNING
Risk of electrical shock.
Do not perform any work on electrical components if the trans-
mitter is switched on. Failure to follow safety instructions could
result in serious injury or death.
Always turn off AC power before manipulating electric parts.
Grounding requirements: Only operate the instrument from
an power outlet which has a ground connection.
Make sure the power specification of the instrument corre-
sponds to the power on site.
Cable
thicknesses
In order to comply with IP66, use the following cable thicknesses
NOTICE: Protect unused cable glands
Wire For Relays: Use max. 1.5 mm
2
/ AWG 14 stranded wire with
end sleeves.
For Signal Outputs and Input: Use 0.25 mm
2
/ AWG 23
stranded wire with end sleeves.
A
B
C
PG 11 cable gland: cable Ø
outer
510 mm
PG 7 cable gland: cable Ø
outer
36.5 mm
PG 9 cable gland: cable Ø
outer
48 mm
ABC