EasyManua.ls Logo

Swann AMI Oxytrace - Carry Out Comparison Measurement

Swann AMI Oxytrace
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AMI Oxytrace QED
Maintenance
A-96.250.531 / 300617 55
6.8.4 Carry out comparison measurement
The comparison measurement is menu driven. Start by selecting
Quality Assurance in menu 3.5 of the monitor AMI Oxytrace.
1 Navigate to menu Maintenance /Quality Assurance.
2 Press [Enter].
3 Follow the dialog on the Display.
4 Carry out pre test preparations
Connect instruments.
Regulate sample flow to 10 l/h us-
ing the appropriate valve.
5 Wait 10 minutes whilst measure-
ment is running.
Press [Enter] to continue.
6 Read the ppb value of the refer-
ence instrument and enter under
“Inspector” by using the [ ] or
[ ] keys.
7 Press [Enter] to confirm.
8 Read temperature value of the ref-
erence instrument and enter un-
der “Inspector Temp.” by using the
[ ] or [ ] keys.
9 Press [Enter] to confirm.
10 Press [Enter] to continue.
The results are saved in QA-
History regardless if successful
or not
3.5.5
Quality Assurance
- Carry out preparations
- Install Inspector
- Sample flow to 10 l/h
<Enter> to continue
3.5.5
Quality Assurance
Value O2 0.05 ppb
Value Temp. 25.00 C
Wait 10 Minutes
<Enter> to continue
3.5.3
Quality Assurance
Value O2 0.05 ppb
Value Temp. 25.00 C
<Enter> to continue
Inspector O2 0.06 ppb
Inspector Temp. 25.0 C
3.5.4
Quality Assurance
Value H2 0.05 ppb
Value Temp. 25.00 C
<Enter> to continue
Inspector 0.06 ppb
Inspector Temp. 25.0 C
3.5.5
Quality Assurance
Max. Dev. O2 0.5 %
Max. Dev. Temp. 0.4 °C
QA-Check succesful
Dev. O2 0.1 %
Dev. Temp. 0.4 °C

Table of Contents

Other manuals for Swann AMI Oxytrace