44 A-96.250.531 / 300617
AMI Oxytrace QED
Maintenance
6.3.2 Clean Swansensor Oxytrace G and Flow Cell
Depending on the water quality, the Swansensor Oxytrace G and
the flow cell will necessitate a cleaning.
Before cleaning, stop operation as described in Stop of Operation
for Maintenance, p. 41.
1 Dismount the Swansensor Oxytrace G, see Electrolyte ex-
change, p. 42.
2 Clean the sensor with a soft tissue and rinse it with water after-
wards.
3 Use a soft brush to remove the dirt which sticks on the walls of
the flow cell.
4 Flush the flow cell with clean water.
5 Install the Swansensor Oxytrace G and start sample flow.