Maintenance
A-96.250.531 / 300617 45
6.4. Maintenance of the Faraday electrode
1 Switch off the instrument and close the flow regulating valve.
2 Open the faraday control unit.
3 Disconnect and remove the cable from the faraday control unit.
4 Unscrew and remove the fixing sleeve (A).
5 Remove the faraday electrode from the flow cell, do not pull on
the cable.
6 Remove the washer (C) and the o-ring [D] from the electrode
body (B).
7 Unscrew the electrode tip containing the hollow electrode (G).
A
B
C
D
E
F
G
Fixing sleeve
Electrode body
Washer
O-ring
Inner electrode
O-ring
Hollow electrode
A
C
D
E
F
G
B