9. Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5 Provides the list of active errors with their status (active, acknowl-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
1.2 Maintenance List
1.2.5 Provides the list of necessary maintenance. Cleared maintenance
messages are moved to the Message list.
1.3 Message List
1.3.1 Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Desig.: Designation of the instrument.
Version: Firmware of instrument (e.g. 6.20-06/16)
2.1.3 Factory Test: Test date of the Instrument and Motherboard.
2.1.4 Operating Time: Shows the operating time in Years, Days, Hours,
Minutes and Seconds.
2.2 Sensors
2.2.1 Oxytrace G
Current value: Shows the actual measuring value in ppb.
Raw value tc: Shows the actual temperature compensated mea-
suring value in mA.
Saturation Shows the actual saturation in %