58 A-96.250.531 / 300617
AMI Oxytrace QED
Maintenance
6.10. Longer Stop of Operation
1 Shut off power of the instrument.
2 Stop sample flow.
3 Remove the Swansensor Oxytrace G.
4 Clean the sensor with a soft tissue and rinse it with water after-
wards.
5 Use a soft brush to remove the dirt which sticks on the walls of
the flow cell.
6 Fill the flow cell with water.
7 Install the Swansensor Oxytrace G.