50 A-96.250.531 / 300617
AMI Oxytrace QED
Maintenance
6.7. Faraday Verification
The Faraday verification works only for oxygen concentrations be-
low 200 ppb. If automatic Faraday verification is enabled, a periodic
check of the system is performed. A manual verification can be
started for test purposes.
Automatic
Verification
Per default the instrument performs an automatic faraday verifica-
tion every 3 hours. To change the settings for automatic verification
go to menu Faraday Parameter- Timer mode, see menu 4.1.3, p.
73 for details.
Possible settings are:
off
interval
daily
weekly
Manual
Verification
To start a manual verification:
1 Navigate to menu 3.2.2 <Maintenance>/<Service>/<Faraday
Verification>.
Results are saved in the Verification history menu 2.2.1.5
2 Press [Enter] to start the Faraday
Verification.
The verification then starts
immediately.
3 Press [Enter] to confirm the Fara-
day Verification.
3.3
Progress
Faraday Verification
<Enter> to stop
Current Value 1.62 ppb
Faraday Conc.
12.85 ppb
3.3
Progress
Faraday Verification
Done
Efficiency 91.5 %
Faraday Conc. 12.85 ppb