72 A-96.250.531 / 300617
AMI Oxytrace QED
Program List and Explanations
3.3 Simulation
In this menu the following relays and signal outputs can be tested:
Alarm relay
Relay 1 and 2
Signal output 1 and 2
Signal output 3 (if option is installed)
Select a relay or signal output with the [ ] or [ ] keys,
press the [Enter]> key to confirm. Then change the value with the
[ ] or [ ] keys. After confirming the setting with the [Enter]
key, the value is simulated by the relay/signal output.
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val-
ues will be reset.
3.4 Set Time
Adjust date and time.
3.5 Quality Assurance
Performs a Quality Assurance according to your settings. Follow
the commands on the screen. Detailed explanation see Quality As-
surance of the Instrument, p. 51.
Alarm Relay: Active or inactive.
Relay 1 and 2: Active or inactive.
Input: Open or closed.
Signal Output 1 and 2: Current in mA
Signal Output 3: Current in mA (if option is installed)