AMI Oxytrace QED
Default Values
A-96.250.531 / 300617 91
Sample Temp., Alarm High: ..................................................... 50 °C
Sample Temp., Alarm Low:........................................................ 0 °C
Alarm Saturation; Alarm high..................................................120 %
Alarm Saturation; Alarm low ....................................................0.0 %
Alarm Saturation; Hysteresis ......................................................2 %
Alarm Saturation; Delay...........................................................5 Sec
Case temp. high: .....................................................................65 °C
Case temp. low:......................................................................... 0 °C
Relay 1 Function:.......................................................................... limit upper
Parameter:............................................................................Oxygen
Setpoint: .......................................................................... 10.00 ppm
Hysteresis:........................................................................... 100 ppb
Delay: ....................................................................................30 Sec
Relay 2 Function:.......................................................................... limit upper
Parameter:....................................................................Temperature
Setpoint: .................................................................................. 50 °C
Hysteresis:.............................................................................. 1.0 °C
Delay: ....................................................................................30 Sec
If Function = Control upw. or dnw:
Parameter:............................................................................Oxygen
Settings: Actuator: .......................................................... Frequency
Settings: Pulse Frequency: .............................................120/min.
Settings: Control Parameters: Setpoint:....................... 10.00 ppm
Settings: Control Parameters: P-band: ......................... 100.0 ppb
Settings: Control Parameters: Reset time:...........................0 Sec
Settings: Control Parameters: Derivative Time: ...................0 Sec
Settings: Control Parameters: Control Timeout:..................0Min.
Settings: Act. Time prop.: Cycle time: ................................... 60 s
Settings: Act. Time prop.: Response time: ............................ 10 s
Settings: Act. Motor valve: Run time: .................................... 60 s
Settings: Act. Motor valve: Neutral zone: ................................5%
If Function = Timer:
Mode:.................................................................................... Interval
Interval: ................................................................................ 1 min
Mode: ........................................................................................daily
Start time:........................................................................ 00.00.00
Mode:..................................................................................... weekly
Calendar; Start time: ....................................................... 00.00.00
Calendar; Monday to Sunday:.................................................. Off
Run time: ...............................................................................10 Sec
Delay: ......................................................................................5 Sec