20 - 36
4.4 Calibrate a Profile Gauge
1. Select the 'Open Calibration' option from the 'File' menu.
2. Select the Profile calibration file from the file selector.
3. Press the 'OK' button.
4. Your file will now be opened.
5. Once the file has finished loading, press 'F2' to begin the program.
6. There will be some on screen prompts and messages during execution of the program,
respond to these as necessary.
7. When the program has completed all operations, the display window will clear.
8. Calibration is now complete.
4.5 Calibration Faults
In the event of a fault during calibration, a system error message or message specific to the gauge
calibration program may be displayed. The gauge calibration program messages are :
There was a problem with the measured length of the setting master.
There was a problem with the measured diameter of the setting
master.
Light may have reached the arrays? i.e. tailstock may have been in
the wrong position.
Light may have been blocked? i.e. tailstock may have been in the
wrong position.
Illumination Dim / Uneven
Defective LED; dirty lenses; faulty array/Cvpb.
There may be an object on the safety paddle switch, E-Stop may be
latched down.
The most likely causes of faults are dirt on the setting master, poor location of the setting master in the
tooling or wrong setting master being used. Check that the serial number on the setting master
corresponds to that displayed on the computer SCREEN, clean the setting master and re-run the
calibration program. If the problem recurs, contact the SYLVAC SA Service organization.
4.6 Measurement Mode
In Measurement mode the measurement system will support the following functions :
Measure a component placed in the gauging device
Display the results of measurement on the SCREEN
Update the statistics (if option available)
When accessing the 'Open Measurement' option after a successful calibration the data display window
will contain a blank screen with the Text display headings. The toolbar measurement control buttons
will become active. After each measurement has been completed, the data will be shown in the table
below their corresponding headings. Each screen full of information is called a page and if the number
of dimensions displayed exceeds the single page limit, the Page control will enable viewing of the next
set of dimensions.