Ifektronix
COMMITTED TO EXCELLENCE
Product: 2205 OSCILLOSCOPE SERVICE
Date: 3-20-90 Change Reference: C2/0390_______
_________________
Manual Part Number: 070-6716-00
MANUAL CHANGE INFORMATION
DESCRIPTION
Product Group 46
EFFECTIVE SERIAL NUMBER: ALL
TEXT CHANGES
PAGE 1-4
CHANGE TO:
Table 1 -1 (cont)
Characteristics
Performance Requirements
Sweep Linearity Magnified
XI X I0
±7 % ’
±10% 2
1 Linearity measured over any 2 of the center 8 divisions.
2 Linearity measured over any 2 of the center 8 divisions, excluding the
first 40 ns.
PAGE 4-6
CHANGE:
Step 1, parts d and h.
d. CHECK-Timing accuracy is within 3% (0.24 division at the tenth vertical graticule line) and linear
ity is within 7% (0.14 division over any 2 of the center 8 divisions).
h. Usethe Horizontal POSITION controls to align the first time marker that is 40 ns beyondthe start of
the sweep with the second vertical graticule line.
PAGE 4 -7
CHANGE:
Step 1, parti.
i. CHECK—Timing accuracy is within 4% (0.32 division at the tenth vertical graticule line) and linear
ity is within 10% (0.2 division over any 2 of the center 8 divisions). Exclude any portion of the
sweep past the 50th magnified division.
PAGE 5-14
CHANGE:
Step 8, parts e, I, and j.
e. CHECK—Timing accuracy is within 3 % (0.24 division at the tenth vertical graticule line) and linear
ity is within 7% (0.14 division over any 2 of the center 8 divisions).
i. Usethe Horizontal POSITION controls to align the first time marker that is 40 ns beyond the start of
the sweep with the second vertical graticule line.
j. CHECK-Timing accuracy is within 4% (0.32 division at the tenth vertical graticule line) and linear
ity is within 10% (0.2 division over any 2 of the center 8 divisions). Exclude any portion of the
sweep past the 50th magnified division.
Page 1 of 1