EasyManuals Logo

Tektronix 2205 Service Manual

Tektronix 2205
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
Adjustment Procedure2205 Service
I. Set the CH 2 VOLTS/DIV switch to 50 mV. 11, Check ADD MODE Operation
m. Repeat parts d through i except adjust the 3
and 3N" trimmers in parts f and i respectively.
n. Set the CH 2 VOLTS/DIV switch to 0.5 V.
o. Repeat parts d through i except adjust the 4"
arid "4N trimmers in parts f and i respectively.
p. Set the Vertical MODS switch to CH 1.
q. Repeat parts b through p for the Channel 1
Attenuators.
r. Disconnect the test equipment from the
instrument.
9. Check Vertical ALT Operation
a. Set:
AC-GND-DC (both)
Vertical MODS
SEC/DIV
Trigger SOURCE
GND
BOTH. NORM
and ALT
0.1 s
CH 1
b. Position the Channel 1 and Channel 2 traces
about two divisions apart using the CH 1 and CH 2
POSITION controls.
c. CHECKChannel 1 and Channel 2 traces
move across the screen alternately.
d. Set Vertical switch to CHOP.
e. CHECKChannel 1 and Channel 2 traces
move across the screen at the same time.
10. Adjust CHOP Switch Balance (R140)
a. Set:
SEC/DIV 1 ms
Trigger MODE NORM
Trigger SOURCE VERT MODE
b. Connect the 10X probe to the rear of R242 on
top of the Al-Main circuit board and adjust test
oscilloscope for a 4-division display.
C. ADJUSTChop-Sw Bal (R140) for no trig
gering on chop segments when rotating the Trigger
LEVEL control.
a. Set:
20 mV
DC
BOTH. NORM,
and ALT
0.5 ms
XI
CH 1
b. Position both traces on the center horizontal
graticule line using the CH 1 and CH 2 POSITION
controls. '
c. Set the calibration generator to produce a
50-mV signal.
d. Connect the output of the calibration generator
to both the CH 1 or X input and the CH 2 o r V input
with dual-input coupler.
e. Check that both channels show a 2.5-division
display.
f. Set: Vertical MODE switch to ADD.
g. CHECKThe resultant display is five divisions
3% (4.85 to 5.15 divisions).
h. Disconnect the test equipment from the
instrument.
12. Adjust High-Frequency Compensation
(C241, C242, C243, and R241)
a. Set:
VOLTS/DIV (both) 10 mV
Vertical MODE C H I
SEC/DIV 0.2 ms
b. Connect the positive-going, fast-rise, square-
wave output from the calibration generator via a
50-0 coaxial cable, a 10X attenuator, and a 50-0
termination to the CH 1 OR X input connector.
c. Set the generator to produce a 1-MHz, 5-
dtvision display.
d. Set the top of the display to the center hori
zontal graticule Tine using the CH 1 POSITION
control.
e. ADJUST-HF Comp (C242 and C243) for the
5% or less overshoot (0.3 division) on the displayed
signal.
VOLTS/DIV (both)
AC-GND-DC (both)
Vertical MODE
SEC/DIV
Horizontal MODE
Trigger SOURCE

Table of Contents

Other manuals for Tektronix 2205

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2205 and is the answer not in the manual?

Tektronix 2205 Specifications

General IconGeneral
BrandTektronix
Model2205
CategoryTest Equipment
LanguageEnglish

Related product manuals