EasyManua.ls Logo

Tektronix 2205 - Check Sweep Holdoff; Check X Bandwidth

Tektronix 2205
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Adjustment Procedure2205 Service
c. Position the trace horizontally so the start of 11. Check Sweep Holdoff
the trace begins at the first vertical graticule line (ex
treme left). a- Set:
d. Set the SEC/DIV switch to X-Y (fully
counterclockwise).
e. ADJUSTX Offset (R736) to position the spot
at the seventh vertical graticule line,
f. Disconnect the test equipment from the
instrument.
VOLTS/DIV (both) I V
AC-GND-DC (both) GND
SEC/DIV 1 ms
Trigger SOURCE EXT, EXT
b. Connect the test oscilloscopes 10X probe tip
to the front end of R704 (toward the front panel).
R704 is on the Timing circuit board.
c. CHECKHoldoff time is between 1.5 ms to
10. Check X Bandwidth 3.0 ms.
50 mV
DC
CH 1
b. Connect the leveled sine-wave generator out
put via a 50-ft coaxial cable and a 50-Xl termination
to the CH 1 o r X input connector.
c. Set the generator to produce an 8-division
horizontal display at an output frequency of 50 kHz.
d. Increase the signal frequency until the hori
zontal deflection (X-axis) is equal to 5.7 divisions in
length.
e. CHECKThe frequency is greater than 2 MHz.
f. Disconnect the test equipment from the
instrument.
d. Repeat part c for SEC/DIV switch settings in
Table 5-5.
e. Disconnect the 10X probe from R704.
Table 5-5
Settings for Holdoff Checks
SiC/OIV
Holdoff Time
0.1 ms
0.2 ms to 0.3 ms
10 us
15 us to 30 us
1 |1 S
1.5 us to 3.0 us
a. Set:
VOLTS/DIV (both)
AC-GND-DC (both)
Trigger SOURCE
5-15

Table of Contents

Other manuals for Tektronix 2205

Related product manuals