Specification—2205 Service
Table 1-1 (cont)
Characteristics Performance Requirements
POWER REQUIREMENTS
Line Voltage Ranges
115V Setting
95 Vac to 128 Vac.®
230 V Setting
185 Vac to 150 Vac.®
Line Frequency
48 Hz to 440 Hz. a
Maximum Power Consumption 40 W (60 VA).a
Line Fuse UL 198.6 3AG (1/4 X 1 1/4 inch)
115 V Setting 0.75 A. Slow.
230 V Setting
0.5 A. Slow.
CATHODE-RAY TUBE
Display Area
8 X 100 mm®
Standard Phosphor
GH (P31).a
Nominal Accelerating Voltage
1800 V ±10%.a
Performance requirement not checked in manual.
1-S