Adjustment Procedure—2215A Service
place instructions located in the Maintenance section of
the manual.
All test equipment items listed in Table 4-1 are required
to accomplish a complete Adjustment Procedure. At the be
ginning of each subsection there is an equipment-required
list showing only the test equipment necessary for perform
ing the steps in that subsection. In this list, the item number
following each piece of equipment corresponds to the item
number listed in Table 4-1.
Before performing this procedure, do not preset any in
ternal adjustments and do not change the 8.6-V power-
supply adjustment. Altering this adjustment may necessitate
a complete readjustment of the instrument, whereas only a
partial adjustment might otherwise be required. Only change
an internal adjustment setting if a Performance Characteris
tic cannot be met with the original setting. If it is necessary
to change the setting of an internal adjustment, check Table
5-1 for possible adjustment interactions.
Before performing any procedure in this section, set the
POWER switch to ON and allow a 20-minute warm-up
period.
The most accurate display adjustments are made with a
stable, well-focused, low-intensity display. Unless otherwise
noted, adjust the INTENSITY, FOCUS, and TRIGGER
LEVEL controls as needed to view the display.
Table 5-1
Adjustment Interactions
A d ju s tm e n ts o r
R ep la c e m e n ts Made
A d ju s tm e n ts A ffe c te d
TRACE ALIGNMENT
GEOMETRY
33NV1V8 00 auis/Z
CH t VAR BALANCE
M F/LF GAIN BAL 4 FREQ COMP
VERTICAL GAIN i
ATTENUATOR COMP
1 DELAY L IN E COMP
1 CH 2 HIGH FREQ MATCH
1 HORIZ GAIN
1 HORIZ X I0 GAIN
1 MAGNIFIER REGISTRATION
1 DELAY D IA L TIM ING
1 HIGH-SPEED TIMIN G
1 5n s TIM ING AND LIN EA R ITY 1
1 TRIGGER OFFSET
1 TRIGGER S E N S IT IV IT Y
1 SLOPE BALANCE
I P -P AUTO TRIGGER CENTERING
-8 .6 V POWER SUPPLY
1
TRACE ALIGNMENT
1
GEOMETRY
2/5m V DC BALANCE
CH 1 VAR BALANCE
|I
M F/LF GAIN BAL 4 FREQ COMP
| I
VERTICAL GAIN
I
ATTENUATOR COMP
| I
DELAY LIN E COMP
m
_
HORIZ GAIN
I _ _ I I
HORIZ X10 GAIN
1 I _
HIGH-SPEED TIM IN G
1 1
_
TRIGGER OFFSET
I I
TRIGGER S E N S IT IV IT Y
I i_
SLOPE BALANCE
_
$
_
CRT REPLACEMENT
L
u
5-2
4206-20

Table of Contents

Other manuals for Tektronix 2215A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2215A and is the answer not in the manual?

Tektronix 2215A Specifications

General IconGeneral
BrandTektronix
Model2215A
CategoryTest Equipment
LanguageEnglish

Related product manuals