EasyManua.ls Logo

Tektronix 2215A

Tektronix 2215A
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Adjustment Procedure—2215A Service
POWER SUPPLY AND CRT DISPLAY
Equipment Required (See Table 4-1):
Leveled Sine-Wave Generator (Item 2) Digital Voltmeter (Item 10)
Time-Mark Generator (Item 3) DC Voltmeter (Item 12)
50-0 BNC Cable (Item 4)
Screwdriver (Item 13)
50-0 BNC Termination (Item 5)
See
ADJUSTMENT LOCATIONSS I
a t the back o f this m anual fo r location o f test points and adjustm ents.
INITIAL CONTROL SETTINGS
Vertical
POSITION (both)
Midrange
VERTICAL MODE
CH 1
CH 1 VOLTS/DIV
10 mV
CH 1 VOLTS/DIV Variable
CAL detent
Channel 1 Input Coupling
GND
srizontal
POSITION
Midrange
HORIZONTAL MODE
A
A SEC/DIV
X-Y
SEC/DIV Variable
CAL detent
X I0 Magnifier
Off (knob in)
TRIGGER
VAR HOLDOFF
NORM
Mode
P-P AUTO
SLOPE
OUT
LEVEL
Midrange
A&B INT
VERT MODE
A SOURCE
INT
PROCEDURE STEPS
1. Check/Adjust Power Supply DC Levels (R938)
a. Connect the digital voltmeter low lead to chassis
ground and connect the volts lead to the -8.6-V supply
(TP961).
b. CHECKVoltmeter reading is 8.56 to -8.64 V. If
the reading is within these limits, skip to part d.
c. ADJUSTThe 8.6 V Adj potentiometer (R938) for a
voltmeter reading of -8.60 V.
d. CHECKVoltage levels of the remaining power sup
plies listed in Table 5-2 are within the specified limits.
Table 5-2
Power Supply Limits
Power
Test
Reading
Supply Point (Volts)
-8 .6 V
TP961
-8.56 to -8.64
+ 5.2 V W968 + 5.04 to +5.36
+ 8.6 V
W960 + 8.43 to +8.77
+ 30 V
W956 +29.1 to +30.9
+ 100 V
W954
+ 97.0 to +103.0
NOTE
Review the information a t the beginning o f the Adjust
m ent Procedure before starting this step.
5-4
e. Disconnect the test equipment from the instrument.

Table of Contents

Other manuals for Tektronix 2215A

Related product manuals