1-4
·Performance Requirement not checked in the manual.
Characteristics
Performance Requirements
AC Coupled Lower -3 dB Point
lX
Probe
10 Hz or less.a
lOX
Probe 1 Hz or less.a
Step Response, Repet and Average On;
Average Set to 16
Rise Time 2.3 ns or less (calculated).s
T, (in ns)
350
BW (in MHz)
Envelope Mode Pulse Response
Minimum Single Pulse Width for 50%
2 ns.a
or Greater Amplitude Capture at
85% or Greater Confidence
Minimum Single Pulse Width for Guaranteed
4 ns.a
50% or Greater Amplitude Capture
Minimum Single Pulse Width for Guaranteed
8 ns.a
80% or Greater Amplitude Capture
Channel Isolation 100:1 or greater attenuation of the deselected
channel
at
100 MHz; 50:1 or greater attenuation at 150 MHz, for a 10-
division input signal from 2 mV/div to 500 mV/div; with equal
VOLTS/DIV settings on both channels.
Acquired Channel 2 Signal Delay with Respect
±250 ps.a
to Channel 1 Signal at Full Bandwidth
Input Rand C (1 MQ)
Resistance
1 MQ ±0.5%.a
In each attenuator, the input resistance of all VOLTS/DIV posi-
tions is matched to within 0.5%.a
Capacitance
15 pF ±2 pF.a
In each attenuator, the input capacitance of all VOLTS/DIV posi-
tions is matched to within 0.5 pF.a
Table
1-1
(cont)
Specification-2430 Service