1-9
·Performance Requirement not checked in the manual.
Sample Rate Accuracy
Average Over 100 or More Samples
±0.01%.a
External Clock
Repetition Rate
Minimum
1 MHz.a
Maximum
100 MHz.a
Events Count 1 to 65,536a
Events Maximum Repetition Rate
100 MHz.a
Signal Levels Required for EXT Clock or EVENTS
Channel 1 or Channel 2 SOURCE
DC Coupled
0.7 division from DC to 20 MHz; increasing to 2.0 divisions at
100 MHz; 3.0 divisions at 100 MHz in ADD mode."
NOISE REJ Coupled
2.4 divisions or less from DC to 20 MHz; increasing to 6.0 divi-
sions at 100 MHz; 9.0 divisions at 100 MHz in ADD mode."
AC Coupled
0.7 division from 60 Hz to 20 MHz; increasing to 2.0 divisions at
100 MHz; 3.0 divisions at 100 MHz in ADD mode. Attenuates sig-
nals below 60 Hz.a
HF REJ Coupled 2.0 divisions from DC to 30 kHz.
Attenuates signals above
30 kHz.a
LF REJ Coupled
2.0 divisions from 80 kHz to 20 MHz; increasing to 4.0 divisions
at 100 MHz; 3.0 divisions at 100 MHz in ADD mode. Attenuates
signals below 80 kHz.a
EXT TRIG 1 or EXT TRIG 2 Source
Ext Gain
=
1
DC Coupled
35 mV from DC to 20 MHz; increasing to 100 mV at 100 MHz.a
NOISE REJ Coupled
120 mV or less from DC to 20 MHz; increasing to 300 mV at
100 MHz.a
AC Coupled 35 mV from 60 Hz to 20 MHz; increasing to 100 mV at 100 MHz.
Attenuates signals below 60 Hz.a
HF REJ Coupled 50 mV from DC to 30 kHz.a
LF REJ Coupled
50
mV from
80 kHz to 20 MHz; increasing to
100 mV
at
100 MHz.a
Ext Gain
=
-i-
5
Amplitudes are five times those specified for Ext Gain
=
1.a
Delay Time Range
(0.04
X
8 SEC/DIV) to (65,536
X
0.04
X
8 SEC/Dlv).a
Delay Time Accuracy
Same as the sample rate accuracy.s
Delay Time Resolution
The greater of (0.04
X
8 SEC/DIV) or 20 ns.
TIME BASE
Performance RequirementsCharacteristics
Table
1-1
(cant)
Specification-2430 Service