1-11
aperformance Requirement not checked in the manual.
CALIBRATOR
CALIBRATOR output amplitudes at 5 MHz are at least 50% of
output amplitudes at 1 ms SEC/DIV settinq."
Voltage (with A SEC/DIV
switch set to 1 ms)
1 Mil Load
O.4V ±1%.a
50 Il Load
0.2 V ± 1.5%.a
Current (short circuit load with
8
mA ±1.5%.a
A SEC/DIV switch set to 1 ms)
Repetition Period A SEC/DIV
Calibrator
Calibrator Div/
Setting8
Frequency8
Period8
Cycle8
5 ns
40
10 ns
20
20 ns
5 MHz 200 ns
10
50 ns
4
100 ns
2
200 ns
1
500 ns
4
1
JLS
500 kHz
2
JL S
2
5
JL S
4
10
JL S
50 kHz
20
JL S
2
20
JL S
1
50
JL S
4
100
JL S
5 kHz 200
JL S
2
200
JL S
1
500
JL S
4
1 ms 500 Hz
2 ms 2
2 ms
1
5 ms
4
10 ms
2
20 ms
1
50 ms
0.4
100 ms
0.2
200 ms 50 Hz
20 ms 0.1
500 ms
0.04
1 s
0.02
2s
0.01
5s
0.004
Accuracy
±0.01%.a
Symmetry
Duration of high portion of output cycle is 50% of output period
± (lesser of 500 ns or 25% of period).a
SIGNAL OUTPUTS
Performance Requirements
Characteristics
Table
1-1
(cont)
Specification-2430 Service