1-12
·Performance Requirement not checked in the manual.
Characteristics Performance Requirements
CH 2 SIGNAL OUTPUT
Output Voltage
20 mV per division ± 10% into 1 Mn.
10 mV per divison ± 10% into 50 n.
Offset
± 10 mV into 50 n, when dc balance has been performed within
± SoC of the operating temperature.
-3 dB Bandwidth DC to greater than 50 MHz.
A TRIGGER, RECORD TRIGGER, and
WORD RECOGNIZER Output
Logic Polarity Negative
true. Trigger
occurrence indicated
by a
HI to LO
transition.a
Output Voltage HI
Load of 400 /-l-Aor Less
2.5 V to 3.5
v»
50 n Load to Ground 0.45 V or greater.a
Output Voltage LO
Load of 4 mA or Less 0.5 V or less.a
50 n Load to Ground
0.15 V or less.a
PLOTTER
X-Output and Y-Output
Output Resistance 1k!1±10%.
Output Range ± (2 V ±100 mV).
Scale Factors
Y-390 mV per division;
X-390 mV per division.
Output Center
o
volts ±30 mV.
Home (Lower Left) Position -2V±100mV.
Slew Rate
Less than 8 volts per secono."
The instantaneous slew rate is determined by the output stage
time constant (3.3 ms) and can exceed 8 volts per second. The
System /-l-Pcomputes the length of the stroke needed for each
point and waits an appropriate time at each position before
proceeding so that the X-Y plotter sees an effective slew rate of
less than 8 volts per second.s
Pen Lift, SPST Relay Contact to Ground
Polarity
Menu selectable.a
Maximum Applied Open-Circuit Voltage
±25 volts."
Maximum Closed-Circuit Resistance
1 nor less.a
Maximum Closed-Circuit Current 250 mA or less.a
Table 1-1 (cont)
Specification-2430 Service