CRoutinetest is with ROD-LIEPAElectronic Model 100AV Hi-Pot Tester. This tests both the Primary Circuit Dielectric Withstand and
Primary Grounding in one operation. Contact Tektronix Product Safety prior to using any other piece of equipment to perform these
tests.
aperformance Requirementnot checked in the manual.
Source Voltage
Nominal Ranges
115 V 90 V to 132
230 V
180 V to 250
v»
Source Frequency 48 Hz to 440 Hz."
Fuse Rating 5 A, 250 V, AGC/3AG, Fast Blow; or 4 A, 250 V, 5 X 20 mm
Time-Lag (T).
a
Each fuse type requires a different fuse cap."
Power Consumption
Typical (standard instrument) 160 watts (250 VA).a
Maximum (fully optioned instrument) 200 watts (300 VA).a
Primary Groundinq? Type test 0.1 S1maximum. Routine test to check grounding con-
tinuity between chassis ground and protective earth qround."
AC POWER SOURCE
Performance Requirements
Characteristics
Table
1-1
(cont)
Specification-2430 Service