OPERATION
45
Output Terminals 2260B-30/80
Background
Before connecting the output terminals to the
load, first consider whether voltage sense will
be used, the gauge of the cable wiring and the
withstand voltage of the cables and load.
The output terminals can be connected to load
cables using M4 sized screws or M8 sized bolts.
WARNING
Dangerous voltages. Ensure that the power to the
instrument is disabled before handling the power
supply output terminals. Failing to do so may lead
to electric shock.
Steps
1. Turn the power switch off.
2. Remove the output terminal cover.
Page 47
3. If necessary, screw the chassis
ground terminal to either the
positive or negative terminal. See
the grounding chapter for details.
Page 36
Sense joining plates
Ground
4. Choose a suitable wire gauge for
the load cables.
Page 44