EasyManua.ls Logo

Tektronix Keithley 2260B Series - Page 89

Tektronix Keithley 2260B Series
207 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPERATION
85
Remote monitoring
Voltage monitoring (VMON) and current
monitoring (IMON) are only supported on the
master unit.
The VMON voltage represents the total voltage
of the all the serialized units.
Remote Sense
Please see the remote sense chapter for details,
page 71.
External Voltage and Resistance Control
Voltage/Resistance controlled remote control
can only be used with the master unit.
Slew Rate
The voltage slew rate is actually double that of
the setting slew rate. I.e., A voltage slew rate
setting of 60.00V/s is actually 120V/s when in
series.
Internal Resistance
The internal resistance is actually twice that of
the setting value.
Only the master unit’s internal resistance
settings apply.
Bleeder Control
The Master unit is used to control the bleeder
settings. The bleeder resistor is always turned
on for the slave unit in series mode.

Table of Contents

Other manuals for Tektronix Keithley 2260B Series

Related product manuals