OPERATION
85
Remote monitoring
Voltage monitoring (VMON) and current
monitoring (IMON) are only supported on the
master unit.
The VMON voltage represents the total voltage
of the all the serialized units.
Remote Sense
Please see the remote sense chapter for details,
page 71.
External Voltage and Resistance Control
Voltage/Resistance controlled remote control
can only be used with the master unit.
Slew Rate
The voltage slew rate is actually double that of
the setting slew rate. I.e., A voltage slew rate
setting of 60.00V/s is actually 120V/s when in
series.
Internal Resistance
The internal resistance is actually twice that of
the setting value.
Only the master unit’s internal resistance
settings apply.
Bleeder Control
The Master unit is used to control the bleeder
settings. The bleeder resistor is always turned
on for the slave unit in series mode.