Commands Listed in Alphabetical Order
Conditions
MDO4000/B/C mo
dels only.
Group
RF
Syntax
RF:RF_FREQuency:VERTical:POSition <NR3>
RF:RF_FREQuency:VERTical:POSition?
Related Commands
RF:RF_FREQuency:VERTical:SCAle
Arguments
<NR3> is a floating point number.
Examples
RF:RF_FREQuency:VERTical:POSition 1 sets the vertical position of the
RF Frequency vs. Time trace to –1 division.
RF:RF_FREQuency:VERTical:POSition? might return 0.0E+0, indicating a
vertical position of 0.
RF:RF_FREQuency:VERTical:SCAle
This command specifies the vertical scale for the RF Frequency vs. Time tr
ace.
For a signal with constant amplitude, increasing the scale causes the waveform to
be displayed smaller. Decreasing the scale causes the waveform to be displayed
larger.
Conditions
MDO4000/B/C models only.
Group
RF
Syntax
RF:RF_FREQuency:VERTical:SCAle <NR3>
RF:RF_FREQuency:VERTical:SCAle?
Related Commands
RF:RF_FREQuency:VERTical:POSition
Arguments
<NR3> is a floating point number that specifies the vertical scale.
Examples
RF:RF_FREQuency:VERTical:SCAle 1.0000E+6 sets the vertical scale to
1 MHz/division.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-577

Table of Contents

Other manuals for Tektronix MDO4000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
BrandTektronix
ModelMDO4000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals