Command Descriptions
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
2- 127
HARmonics:PHAse?
Might return 45.0, indicating that the phase of the select ed
harmonic, relative to the fundam ental frequency, is 45 degrees.
HARmonics:SELect
HARmonics:RMS? (Query Only)
(TPS2000 Series with TPS2PWR1 Power Analysis Module Only)
Returns the amplitude of the harmonics source in RMS. This may be
V
rms
or I
rms
depending on the type of source wave form.
Power Measurement
HARmonics:RMS?
HARmonics:RMS?
Might return 120.0, indicating that the RMS value is 120 volts
RMS.
HARmonics:SAVe (Set Only )
TPS2000 Series with TPS2PWR1 Power Analysis Module Only)
Sets the file name a nd pat h to save the selected waveform harmonic
data to the oscilloscope CompactFlash card in .CSV format.
The following data is included in the file:
H Harmonics, 1 through 50
H Magnitude
H Percent of fundamental
Examples
Related Commands
Group
Syntax
Examples