Command Descriptions
2- 128
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
H Frequency
H Phase angle
NOTE. You can view the contents of the saved file on your personal
computer. You cannot recall saved harmonics files to the oscillo-
scope.
Power Measurement
HARmonics:SAVe <file path>
<file path> specifies the full path to save harmonic measurement
data.
HARmonics:”SAVe A:\data1.CSV”
Specifies the file name and path to save harmonic loss measurement
data to the oscilloscope CompactFlash card.
HARmonics:SELect
(TPS2000 Series with TPS2PWR1 Power Analysis Module Only)
Selects an individual harmonic. If the harmonics function is enabled
but the requested harmonic is not currently on the screen, the
oscilloscope will put it on the screen. If the harmonics function is not
enabled, the oscilloscope will remember the requested value but will
not enable the ha rmonics function.
Power Measurement
Group
Syntax
Arguments
Examples
Group