Command Descriptions
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
2- 139
HORizontal:MAIn:SECdiv
Sets the time per division for the oscilloscope main time base. This
command is identical to the HORizontal:MAIn:SCAle command. It
is provided to maintain program compatibility with some older
models of Tektronix oscilloscopes.
HORizontal:POSition
Sets or queries the main time base horizontal position. This
command is identical to the HORizontal:MAIn:POSition command.
This command is included for compatibility.
Horizontal
HORizontal:POSition <NR3>
HORizontal:POSition?
<NR3> is the position in seconds. This value is the difference between
the trigger point and the center graticule. Positive values place the
trigger before the center graticule.
HORizontal:POSition 2.0E–6
Sets the main trigger position to 2 s before the center graticule.
HORizontal:POSition?
Might return -1.0E-3, indicating that the main trigger position is
1 ms after the center graticule.
Group
Syntax
Arguments
Examples