Command Descriptions
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
2-77
VOLTSSQUARED indicates volts squared (V*V) as the unit of measure.
(TDS1000B, TDS2000B, and TPS2000 series only)
AMPSSQUARED indicates amps squared (A*A) as the unit of measure.
(TDS1000B, TDS2000B, and TPS2000 series only)
VOLTSAMPS indicates voltage times current (V*A) as the unit of
measure. (TDS1000B, TDS2000B, and TPS2000 series only)
NOTE. Unknown units are represented by ā?ā in the oscilloscope
readouts.
CURSor:HBArs:UNIts?
Might return CURSOR:HBARS:UNITS VOLTS.
CURSor:SELect:SOUrce
Sets or queries the waveform that is the source of the vertical and
horizontal scale factors used in determining cursor values. This
command is equivalent to setting the Source option in the Cursor
menu.
Cursor
CURSor:SELect:SOUrce <wfm>
CURSor:SELect:SOUrce?
<wfm> specifies the waveform data source on which cursor
measurements will be taken.
Examples
Group
Syntax
Arguments