Command Descriptions
2-76
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
NOTE. The source always determines the measurement units.
CURSor:HBArs:POSITION1 25.0E–3
Positions one of the horizontal cursors at 25.0 mV (assuming the
vertical units are volts).
CURSor:HBArs:POSITION2?
Might return –6.40E–2, indicating that the second horizontal bar
cursor is at –64.0 mV (assumi ng the vertica l units are volts).
CURSor:HBArs:UNIts? (Query Only)
Returns the vertical scale units for the selected cursor source
waveform.
Cursor
CURSor:HBArs:UNIts?
VOLts indicates volts from ground as the unit of measure.
DIVs indicates divisions as the unit of measure, with center of screen
as 0 divisions and bottom of screen as –4 divisions.
DECIBELS indicates decibels as the unit of measure, relative to a
1V
rms
sine wave. (FFT only)
UNKNOWN indicates that Trigger View is active. This also generates
event message 221. (Settings conflict)
AMPS indicates amps as the unit of measure. (TDS1000B,
TDS2000B, and TPS2000 series only)
Examples
Group
Syntax
Returns