Command Descriptions
2- 188
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
OFF turns off pass/fail beep on failure.
ON turns on pass/fail beep on failure.
<NR1> is an integer number where NR1 = 0 turns off pass/fail beep on failure,
and NR1 0 turns on pass/fail beep on failure.
MASK:TEST:BEEP:FAILURE OFF
turns off pass/fail beep on failure.
MASK:TESt:DELay
(Requires TDS3TMT)
Sets or returns the amount of time, in seconds, to wait after starting pass/fail
testing to evaluate waveforms. This is useful if the test system requires ‘settling’
time prior to evaluating waveforms.
Mask
MASK:TESt:DELay <NR3>
MASK:TESt:DELay?
<NR3> is a floating point number that specifies the amount of time, in seconds, to
delay the start of the pass/fail test.
MASK:TEST:DELAY 1
delays waveform evaluation by 1 second after the start of pass/fail testing.
MASK:TESt:HARDCopy
(Requires TDS3TMT)
Sets or returns the hard copy on pass/fail test failure mode. When enabled, this
command causes the oscilloscope to generate a hard copy, using the current
oscilloscope hard copy settings, as soon as the pass/fail testing status changes to
‘Failing.’