Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 189
NOTE. The oscilloscope generates a hard copy only on the first violation that
causes the pass/fail test to fail. To generate a hard copy file for every violation
occurrence, set MASK:TESt:REPeat to ON and set MASK:TESt:WAVEform to 1.
Use this technique only when the expected number of violations is small and the
violation occurs intermittently.
Mask
MASK:TESt:HARDCopy { OFF | ON | <NR1> }
MASK:TESt:HARDCopy?
OFF turns off generate hard copy on failure.
ON turns on generate hard copy on failure.
<NR1> is an integer number where NR1 = 0 turns off generate hard copy on
failure, and NR1 0 turns on generate hard copy on failure.
MASK:TEST:HARDCOPY ON
turns on generate hard copy on failure.
MASK:TESt:REPeat
(Requires TDS3TMT)
Sets or returns the repeat pass/fail test on completion mode. When enabled, this
command repeats the pass/fail test cycle at the completion of the previous test
cycle.
Mask
MASK:TESt:REPeat { OFF | ON | <NR1> }
MASK:TESt:REPeat?
OFF turns off repeat on completion.
ON turns on repeat on completion.
<NR1> is an integer number where NR1 = 0 turns off repeat on completion, and
NR1 0 turns on repeat on completion.
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments

Table of Contents

Other manuals for Tektronix TDS3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3000 Series and is the answer not in the manual?

Tektronix TDS3000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals