Command Descriptions
2- 208
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
MEASUrement:IMMed:DATa? (Query Only)
Returns the measurement result and qualifier for the immediate measurement
specified by the MEASUrement:IMMed:TYPe command. The measurement is
taken on the source specified by a MEASUrement:IMMed:SOURCE command.
Returns both the measurement result and the status of the result.
Measurement
MEASUrement:IMMed:DATa? <NR3>, <NR1>
<NR3> is the measurement result.
<NR1> is one of the execution errors shown in Table 3 --5 on page 3--14, or it is 0
if there is no error.
MEASUREMENT:IMMED:DATA?
might return :MEASUREMENT:IMMED:DATA 9.91963315E-4, 0
MEASUrement:IMMed:VALue?
MEASUrement:IMMed:DELay? (Query Only)
Returns information about the immediate delay measurement.
Measurement
MEASUrement:IMMed:DELay?
MEASUREMENT:IMMED:DELAY?
might return :MEASUREMENT:IMMED:DELAY:EDGE1 RISE;EDGE2 RISE;
DIRECTION FORWARDS