Tektronix WFM8200 - Figure 51: Sample of Transfer.exe Window after the Upgrade Is Complete

Tektronix WFM8200
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Functions
5. Perform the sof
tware upgrade:
a. On the PC, double-click the transfer.exe le to launch the transfer
program.
b. Type the IP address or DNS name of the instrument you are upgrading,
andthenpre
ss Enter. This starts the software upgrade process.
CAUTION. To prevent upgrade failure, do not close the transfer.exe window
until the transfer utility asks for another IP address.
When the software upgrade completes, the instrument may reboot.
c. The transfer utility displays done when the software upgrade completes.
When the software upgrade completes, the instrument will reboot.
Figure 51: Sample of transfer.exe window after the upgrade is complete
6. If the instrument did not reboot at the completion of the software upgrade,
rem
ove and then reapply power to the instrument to cause it to reboot.
Verify Software
1. If the instrument did not reboot at the completion of the software upgrade,
cycle the power to the instrument to cause it to reboot.
2. Press the CONFIG button on the front panel.
3. Navigate to Utilities > View HW/SW Version.
4. Verify that the software version number matches the version of the
software-upgrade package you installed.
100 WFM8200 and WFM8300 Waveform Monitors