Monitoring the SDI Physical Layer
Figure 54: Measuring rise time using variable gain in eye
To Make Automatic Eye
Measurements
NOTE. If you use the measurement cursors to ma nually measure the eye
waveform, you might get a different result than that shown in the automatic eye
measurements readout. This is because the automatic amplitude measurement
uses histograms to be minimally affected by overshoot, ringing, and noise on
the signal. S imilarly, the automatic rise time and fall time measurements use
histo
grams to nd the centers of the distributions of 20% and 80% crossings. The
difference between manual and automatic measurements is usually insignicant
for clean, symmetrical signals.
1. Perform the initial setup for Eye measurements. (See page 105, Eye
Measurement Set Up.)
2. Press and hold the STATUS button to display the pop-up menu.
3. Select Dis
play Type and then select SDI Status.
4. Note the Eye waveform measurements in the SDI Status display readouts.
Automati
c Eye measurements (and a histogram of the Eye waveform) are
also displayed in the Eye Waveform display when expanded to the single,
full-screen display.
108 WFM8200 and WFM8300 Waveform Monitors