Getting Started
3. Press the Power
button on the front panel to turn the instrument on. It is
located o n the bottom left corner of the front panel.
Figure 3: Front panel
4. After about 30 seconds, the power-on diagnostic page appears on the screen.
5. Verify that all self tests pass. Any failures are shown in red. The results of
the power-on diagnostics are erased from the screen, but you can view the
results by selecting CONFIG > Utilities > View Diagnostics Log > SEL or
by viewing them on the instrument Web page.
6. After the diagnostics are nished, the instrument state is restored. When the
progress indicator in the status bar is nished, the instrument has nished
initializing.
7. Record Pass or Fail for the POST (Power On Self Test) in the test record.
8. If it is still open, exit the Diagnostics Log.
Front Panel Test
1. Press and hold the PRESET button.
2. Use the GENERAL knob to navigate to Recall Preset.
3. Press the right and down arrow keys to navigate to Recall Factory Preset and
press the SEL button.
4. Press the FULL button to change the display screen from four-tile view to
single-tile view.
5. Press the HELP buttontodisplaythehelpscreen.
12 WFM8200 and WFM8300 Waveform Monitors