Chapter 4 – Temperature Controller
Page 4-4 TestEquity 1007C Temperature Chamber
System Enable Function
The chamber can be configured to enable or disable all chamber functions through Event 1
(Digital Output 1) in the F4 Controller. The CONDITIONING Switch needs to be in the Event 1
position. This configuration may be desirable if you want to turn off all chamber function at the
end of a programmed profile, or through the communications interface.
All chamber functions will be disabled if Event 1 is Off. If Event 1 is On, all chamber functions
will be enabled, as long as the Power Switch is also On.
NOTE: The Modbus register to program Digital Output 1 in static set point mode is 2000.
NOTE: In the Profile Create and Edit menus, Digital Output 1 is named CONDITION.
Remember to set CONDITION to On in all programmed steps when the chamber is to be
operational.
To Enable or Disable all chamber functions in static set point mode:
1. The F4 Temperature Controller must first be on the Main
Page. Press the ▲or ▼ key to position the Cursor next to
the DigitalOut prompt.
SP1 85.0°C ▲
► DigitalOut_______
█
Power1 50% ▼
2. Press the ► key once. You will see
Choose Event Output prompt in the lower
display. Press the ▲or ▼ key to position the Cursor
next to Event Output1.
► Event Output1
Event Output2
█
Event Output3 ▼
3. Press the ► key once to select Event Output1.
Then, press the ▲or ▼ key to select On or Off.
Off
► On
4. Press the ► key once to return to the
Choose Event Output prompt.
5. Press the ◄ key once to return back to the Main Page.
You will see a 1 in the DigitalOut line if Event1 is
On.
SP1 85.0°C ▲
► DigitalOut1______
█
Power1 50% ▼