Chapter 8 Specifications
Page 8-2 TestEquity 1007C Temperature Chamber
F4 Temperature Controller Specifications
Specifications as configured for the TestEquity 1007C Chamber
Accuracy & Sensor Conformity* ±1.55°C (above 50°C)
±1.66°C (below 50°C)
Stability ±0.1°C/°C rise in ambient
Digital Inputs (Four) Contact closure or dc voltage, 10 k impedance
Retransmit Outputs (Optional) (Two) User-selectable ranges:
0 to 10 VDC, 0 to 5 VDC, 1 to 5 VDC
0 to 20 mA, 4 to 20 mA
Alarm Outputs (Two) Electromechanical relay;
Form C, 2 A @ 20 VDC or 240 VAC max.
Digital Outputs (7 available for customer use) Open collector output
OFF: 42 VDC @ 10 µA max.
ON: 0.2 VDC @ 50 mA sink max.
Internal supply: 5 VDC @ 80 mA
Communications EIA-232 and EIA-485 serial communications with
Modbus™ RTU protocol
Safety & Agency Approvals UL/c-UL 916-listed, File #E185611
CE to EN61010
NEMA 4X and IP65
CE EMC to EN50082-2
CE EMC to EN55011
Displays Process: 5, seven-segment red LED.
Interface Display: 4-line high-definition green LCD;
selectable °C or °F
Data Retention Retention upon power failure via nonvolatile memory
(seven years for battery-backed RAM)
*Note: Total system accuracy in the chamber includes thermocouple wire accuracy.
Thermocouple wire accuracy is ±1°C or 0.75% of reading, whichever is greater. Therefore, total
system accuracy over the chamber’s operating range can be as much as ±2.66°C, although the
typical accuracy is often better than ±1.0°C.