Chapter 4 – Temperature Controller
Page 4-6 TestEquity 1007C Temperature Chamber
Soak Step
Soak maintains the set point from the previous step for a chosen time in hours, minutes and
seconds. Define the Soak step by choosing:
• *Wait for an event or process value
• Event outputs 1 through 7 to turn ON or OFF (For controlling the power to remote devices.)
• Time
• PID set (One of five sets of PID tuning parameters. Normally, just leave at PID Set 1.)
• **Guaranteed Soak
Jump Step
Jump initiates another step or profile. Define the Jump step by choosing:
• Profile to jump to;
• Step to jump to; and
• Number of Repeats
NOTE: If a power-out condition occurs during a profile and more than 20 jump steps are stored
in the F4’s Profile Program memory, the controller will terminate the profile and turn off all
outputs if Continue, Hold or Terminate was selected as the Power Out action. If Profile Reset or
Go to Idle Set Point was selected, the controller will take those actions. A pop-up message will
warn of this when the 21st jump step is programmed
End Step
End terminates the profile in a chosen state. All profiles must have an End step. It cannot be
deleted or changed to another step type. Define the End by choosing Hold, Control Off, All Off
or Idle end state.
NOTE: TestEquity recommends having the end step type to be Hold or Idle. TestEquity does
NOT recommend using an end step type of Control Off or All Off. This does not turn off the
chamber fan. The chamber temperature can reach +55°C just from heat generated by the fan and
interior light, or even higher if your test sample is energized.
*Wait For step option
The use of Wait For in a profile is optional. Ramp Time, Ramp Rate and Soak steps can be
programmed to wait for a particular chamber temperature or event input condition. The wait
conditions must be satisfied before the time clock and the step activity proceeds. Digital inputs
must first be configured in the System Menu as Wait For Events, with the condition to be met
also specified. Then, to wait for this digital input, you must specify On, meaning the condition as
configured in the Setup Page, or Off, meaning the opposite of that condition. The digital inputs
have been configured to Off by TestEquity, so this option will not show unless reconfigured.
**Guaranteed Soak step option
The use of Guaranteed Soak in a profile is optional. The Guaranteed Soak step requires the
chamber temperature to be at the set point temperature, within the Guaranteed Soak Band value,
before the time clock and the step activity proceeds. The Guaranteed Soak Band is configured by
TestEquity for 3.0°C, and this can be changed in the System Menu.