8 Using the product
33
2
Remove particle filter (1) and store it
to reinsert it after the measurement.
3
Remove the sealing plug (2) in the filter chamber from the holder.
4
Close the gas path with the sealing
plug.
5
Check that the sealing plug is fitted tightly. It should not yield at all on
being tugged gently.
6
Close filter chamber of the flue gas probe.
CAUTION
Hot probe shaft! Risk of burns!
- Allow the probe shaft to cool down after a measurement, before
touching it!
- Only attach the silicone hose to the probe shaft once it has cooled
down!
7
Fit silicone hose onto the probe shaft
of the flue gas probe. The probe
shaft openings must be closed.