5 Using the product
25
5.3.1.2. Using the flue gas probe
Checking the thermocouple
The thermocouple of the flue gas probe must not lie against the
probe cage.
> Check before use. Bend the thermocouple back if necessary.
Aligning the flue gas probe
The flue gas must be able to flow freely past the thermocouple.
> Align the probe by turning it as required.
The tip of the probe must be in the core current of the flue gas flow.
> Align the flue gas probe in the flue gas duct so that the tip is in
the core current (area of the highest flue gas temperature).
5.3.1.3. Setting fuel
To carry out a flue gas measurement, the fuel must be set
correctly, see Measurements configuration menu, page 23.
Pos: 47 /TD/Produk t verwenden/testo 310/ testo 310 Abgasmessung @ 12\ mod_1332427125680_79. docx @ 114365 @ 331155555555 54554 @ 1