3 Specifications
9
Measure-
ment
parameter
Measurement
range
Resolution Accuracy Response time
t90
COamb 0 to 4000 ppm 1 ppm ± 3 ppm (<30 ppm)
± 10% of meas. val. (31 to
200 ppm)
± 20 ppm (201 to 400 ppm)
± 5% of meas. val. (401 to
2000 ppm)
± 10% of meas. val. 2001 to
4000 ppm
60s
Draught -20.00 to 20.00
hPa
0.01 hPa ± 0.03hPa (-3.00 to 3.00 hPa)
±1.5% of meas. val. (rest of range)
-
∆P
-40.0 to 40.0 hPa 0.1 hPa ± 0.5hPa
-
Flue gas
temperature
0.0 to 400.0 °C 0.1°C
± 1°C (0.0 to 100.0°C)
± 1.5% of meas. val. (>100°C)
<50s
Ambient
temperature
-20.0 to 100.0 °C 0.1°C
± 1°C
<50s
Pos: 30 /TD/Leis tungsbeschreibung/ Technische Daten/t esto 310 Technische Daten/ weitere Gerätedat en testo 310 @ 12\mod_133240916 6268_79.docx @ 113855 @ 35 @ 1
3.2.2. Other instrument data
Flue gas analyser
Feature Values
Storage and
transport
temperature
-20 to 50 °C
Operating
temperature
-5 to 45 °C
Power supply
Rechargeable battery: 1500mAh
Mains unit: 5 V/1 A
Protection class IP40
Weight incl. probe approx. 700g
Dimensions 201 x 83 x 44 mm
Battery charge time Approx. 5-6 hours
Rechargeable
battery life
> 8h (pump on, 20°C ambient temperature)
EU Directive 2004/108/EC