6 Maintaining the product
31
2. Open the sealing plug of the condensate trap.
3. Let the condensate run out into a sink.
4. Dab off any remaining drops on the condensate outlet with a
cloth.
5. Close condensate outlet with sealing plug and press on it firmly.
The condensate outlet must be completely closed, other-
wise measuring errors could occur if external air gets in.
Pos: 55 /TD/Produk t instand halten/tes to 310/testo 310 Instandh altung Partikelfi lter prüfen @ 12\mod_133 2429814300_79.doc x @ 114570 @ 12225555 @ 1
6.4. Checking / replacing the particle filter
Checking the particle filter:
> Check the particle filter of the flue gas probe for contamination
at regular intervals: check visually by looking through the
window of the filter chamber.
Replace the filter if there are signs of contamination.
Replacing the particle filter:
The filter chamber may contain condensate.
1. Open the filter chamber: Turn gently anti-clockwise.
2. Remove the filter and replace it with a new one (0554 0040).
3. Attach the filter chamber and lock it Turn gently clockwise.