5 Using the product
26
5.3.2. Flue gas
Select measurement type
> Select : [ ] → [OK].
Carry out the measurement
1. Start measurement: [Start].
- The readings are displayed.
> Change reading display line 1: [▲].
> Change reading display line 2: [▼].
2. Quit measurement: [Stop].
> Remove flue gas probe from the flue gas duct and purge with
fresh air.
Pos: 48 /TD/Produk t verwenden/testo 310/ testo 310 CO-Umgebung @ 12\ mod_1332428458095_7 9.docx @ 114399 @ 32325555555 5 @ 1
5.3.3. Ambient CO
Cigarette smoke influences the measurement by more than
50 ppm. The breath of a smoker influences the
measurement by about 5 ppm.
The probe must be in the open air (CO-free) during the
zeroing phase!
Select measurement type
> Select : [ ] → [OK].
Carry out the measurement
1. Start measurement: [Start].
- The reading is displayed.
2. Quit measurement: [Stop].
Pos: 49 /TD/Produk t verwenden/testo 310/ testo 310_Zugmessung @ 12\ mod_1332428743903_7 9.docx @ 114433 @ 2112555555 @ 1
3.
5.3.4. Draught measurement
Do not measure for longer than 5 min, as a drift of the
pressure sensor means that the readings could be outside
the tolerance limits.
Select measurement type
> Select
: [ ] → [OK].