7 Tips and assistance
32
Pos: 56 /TD/Übersc hriften/8. Tipps und Hilfe @ 0\mod_117378988 7985_79.docx @ 406 @ 21555555 5 @ 1
7 Tips and assistance
Pos: 57 /TD/Übersc hriften/8.1 Fragen und An tworten @ 0\mod_117740 2017078_79.docx @ 1093 @ 2212 55 @ 1
7.1. Questions and answers
Pos: 58 /TD/Tipps u nd Hilfe/Fragen und Antwor ten/testo 310 Fragen und An worten @ 12\mod_1332430 291744_79.doc x @ 114604 @ 22555555 @ 1
Question Possible causes / solution
Rechargeable battery low > Switch to mains operation.
Measuring instrument
switches off automatically
or cannot be switched on
Batteries / rechargeable batteries
empty.
> Charge rechargeable battery or
switch to mains operation.
When switching off, the
instrument rinses the gas
path for a long time and
does not shut down.
The gas path is closed by the sealing
plug.
> Remove the sealing plug and
insert the filter.
Error message: E04
O2 sensor is worn out
> Contact Testo Service
Error message: E05 O2 measuring value is outside the
measuring range
> Note the measuring range (see
Technical data)
Error message: E06
Zeroing in the flue gas duct
> Carry out zeroing outside the flue
gas duct (zeroing is repeated a
maximum four times).
Error message: E08 CO measuring value outside the
measuring range (>4000ppm)
> Note the measuring range (see
Technical data)
> Purge CO sensor with fresh air to
prevent destruction of the CO
sensor.
Error message: E12 / E13
CO measuring value unstable
> Carry out zeroing outside the flue
gas duct (zeroing is repeated a
maximum four times).
Error message: E14
Checksum error
> Critical error, contact Testo
Service