6 Using the product
26
Symbol Explanation
Probe info Probe name, serial number and probe
type are displayed.
Probe name Probe name can be edited.
Adjustment info The probe-
specific adjustment data saved
on the probe can be displayed.
The digital probes allow direct
measurement and signal
conversion in the probe. This
technology means that
measurement uncertainty caused
by the instrument is no longer an
issue.
The probe can be calibrated
without a hand instrument.
Entering the adjustment data via
the EasyClimate software
generates a zero-error display.
Humidity
adjustment
Humidity adjustment is possible with the
following probes:
Humidity probe
IAQ probe
Thermal air flow measuring head and
humidity probe
When a thermal air flow
measuring head and humidity
probe is used, the probe must be
deactivated before the
adjustment.
Probe reset
Probe is reset to the factory settings:
Measurement value display
Probe name
Adjustment table
Humidity adjustment
Damping