7 Maintaining the product
55
The Testo adjustment set provides a reference value to calculate
the offset for a humidity adjustment.
Humidity adjustment is possible with the following probes:
Humidity probe
IAQ probe
Thermal air flow measuring head and humidity probe
When a thermal air flow measuring head and humidity
probe is used, the probe must be deactivated before the
adjustment ([ ] Switch off flow probe). The probe
can only then be exposed to the reference conditions.
Instrument is switched on and probe is connected.
Probe is already exposed to reference conditions (e.g. salt pot)
for an adequate adjustment time.
Adjustment time for humidity probe at least 30 min
Adjustment time for IAQ probe at least 1 h
Adjustment time for thermal air flow measuring head and
humidity probe at least 3 h
Tab for the relevant probe is activated.
1. [ ] Probe menu Humidity adjustment.
2. Select reference value used (11.3%RH/75.3%RH).
3. [ ] Adjustment.
- A pop-up window with the remaining adjustment time is
displayed.
4. [ ] End.
- Adjustment menu is closed.
5. Carry out adjustment for the second adjustment point.
7.1.3. Carrying out a firmware update
Updated firmware can be installed on the instrument.
A firmware update can also be carried out with testo
EasyClimate.
Instrument is switched on.
1. Download the current firmware file from
www.testo.com/download-center (registration required).
2. Unzip downloaded zip file.
3. Connect the instrument to the PC with a mini USB cable.
- Instrument is identified by the PC as a USB mass storage
device. The operating system automatically assigns the