6 Using the product
22
Symbol Explanation
Factory reset The instrument is reset to the factory
settings:
Settings
Adjustment data
Password deactivated
Favourites are deleted from the
internal tab. Favourites remain on the
probe tab.
3. Exit the configuration menu: [ESC] [ESC]
- Instrument switches to the measuring view.
6.2. Setting the measurement display
The measurement display can be set individually for each probe
tab. These settings are saved in the probe and also apply when the
probe is next connected.
1. Select the tab for which the measurement display is to be
changed: press [ ].
2. Press [ ].
- The Options configuration menu is opened.
3. Press [ ] to select the Display menu.
4. Open the Display menu: press [ ].
- The Display menu contains three different sub-menus
Selectable sub-menus
Representation Explanation
Min/Max
When the function is enabled, each
line in the measuring view shows: mean
value, min. value and max. value.
Disable function: [ ] [ ] Display
Min/Max [ ].