6 Using the product
31
6.6.3. Measuring program
You can customise measuring programs to match the relevant
measuring task (e.g. timed or multi-point storage). These
measuring programs are linked to a specific measuring point. After
the measurement, the relevant measurement protocols are saved
under the measuring program.
For creating a new measuring program,
A measuring program is always created under a measuring point.
1. Press [ ].
- Explorer structure is displayed.
2. Select measuring point: [ ] and confirm with [ ].
3. [ ] → New measuring program.
4. Determine parameters.
Parameter Explanation
Name Name under which the measuring program is
saved in Explorer.
Measuring point Measuring point to which the measuring
program is assigned.
Measuring type
• Continuous: Mean calculation across a
specific period
• Punctual: Mean calculation across
individual current values buffered via [ ]
or button on probe handle.
• Continuous/Punctual: At each point a
mean value is calculated according to the
end criterion set (Duration or Number of
values). At the end of the measurement,
an overall mean value is calculated.
Measuring rate
Interval at which the measurement values are
recorded.
Start criterion
• Manual: Measurement is triggered via [
].
On the probe handle there is a
button, which can also be used to
start the measurement.
• Date/Time: Measurement starts at a set
time.