Thermo Scientific APEX User’s Guide Page 151
Calibrating the Phase-Angle Setting
The phase angle (and detect level and AuditCheck settings) of the product is learned by
the detector during its auto-calibration procedure. For auto-calibration procedures for
your specific application, consult the list below.
• Conveyor applications—see page 55.
• Gravity-feed applications—see page 73.
• Pipeline applications—see page 90.
• Pharmaceutical applications—see page 105.
Recalibrating the Phase Angle
This calibration menu allows you to recalibrate the phase angle setting by running
product through the search head.
To recalibrate the phase angle, do the following
1) Make sure the calibration function is highlighted.
Calibrate Phase Angle
2) Press the Go button and a calibrate monitoring screen appears.
Pass Product to Calibrate Phase Angle
3) Pass the product through the search head. When the green indicator light in the
control panel flashes or stays on, the phase-angle setting in the screen above has
been updated.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific APEX 100 and is the answer not in the manual?

Thermo Scientific APEX 100 Specifications

General IconGeneral
BrandThermo Scientific
ModelAPEX 100
CategoryMetal Detector
LanguageEnglish

Related product manuals